| Model Number |
GS2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pressure Sore/Ulcer (2326)
|
| Date of Event |
09/10/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient was admitted on (b)(6) 2022.A dtpi was observed in the sacrum/natal cleft area on (b)(6) 2022.No further information on treatment of the dtpi was provided.Evaluation of guardian data noted several gaps in therapy, including a total of 19 hours 5 mins due to enhancer disconnect.This included 4 episodes, totaling 8hrs 22 mins between (b)(6) 2022.It was also noted the incorrect patient weight range was selected (selected range was 126-136kg).There were no indications of device malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted based on a retrospective review (reference rca # (b)(4).Investigation was conducted on the reported information and from the data collected by the guardian device available at the time.
|
| |
|