| Model Number |
1845010 |
| Medical Device Problem Codes |
Overheating of Device (1437); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis completed that, not able to confirm reported fault.The temperature test was not able to perform as it was difficult to insert the tool into the attachment.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
A health care professional reported that pre-op, the indigo attachment was heating and the tool was stuck.There was no patient involved in the event.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 : additional information received.H3 : analysis of the device was completed and found that the bearings are detached and also corroded.H6 : coding updated.Previously applied code fdc d02 is no longer applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received that the issue occured during testing and overheating was noticed within a minute.The device speed was at 36000 rpm, was being run in forward mode.No irrigation was used.
|
| |
|
Search Alerts/Recalls
|