| Model Number |
N/A |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the catheter lumen with a blue port of a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set was noted to be occluded immediately after placement.The catheter was replaced during the same procedure.As reported, there were no adverse effects or additional procedures for the patient due to this occurrence.Additional information regarding the event has been requested but is currently unavailable.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.D1: cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set.E1- customer (person): (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfg.= other (81) - device evaluation has begun; however, a conclusion is not yet available.Device has been returned and preliminary evaluation has been performed, however, our investigation is ongoing and device evaluation summary will be included in our follow up report once our investigation has been completed.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
It was reported in additional information that the lumens were filled with either saline solution or heparinized saline solution prior to catheter introduction.All the lumens were able to be flushed successfully prior to placement.All unused lumens were maintained with either continuous saline, heparinized saline drip, or locked with heparinized saline solution.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation ¿ evaluation.It was reported that the catheter lumen with a blue port of a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set was noted to be occluded immediately after placement.The catheter was replaced during the same procedure.There were no adverse effects or additional procedures for the patient due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions (mi), and quality control procedures, as well as a visual inspection and functional test of the returned device, were conducted during the investigation.Cook received one used catheter, multiple needles, two 8.5 dilators, a movable suture wing, and a wire guide holder.During tabletop testing, all lumens were open; there was no damage noted to any of the components.Cook was unable to replicate the failure and did not identify any non-conforming material.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot and related subassembly lots revealed no recorded non-conformances relevant to the failure mode.A complaint history search identified one other event associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The product ifu, [c_t_ctulmabrm_rev7] ¿cook spectrum central venous catheter minocycline/rifampin antibiotic impregnated power injectable,¿ provides the following information to the user related to the reported failure mode: warnings ¿failure to ensure patency of the catheter prior to power injection studies may result in catheter failure.¿ clincal studies product recommendations suggested catheter maintenance ¿to prevent clotting or possibility of air embolus, the double-lumen¿s #2 lumen, the triple-lumen¿s #2 and #3 lumens and the five-lumen¿s #2, #3, #4, and #5 lumens should be filled with saline solution or heparinized saline solution (100 units of heparin per ml of saline is usually adequate), depending on institutional protocol, prior to catheter introduction.¿ instructions for use ¿2.Prepare the catheter for insertion by flushing each of the lumens and clamping or attaching the injection caps to the appropriate extensions.¿9.-lumens should now be flushed with 5-10ml normal saline prior to use or establishment of heparin lock.¿ based on the information provided, inspection of the returned device, and the results of the investigation, cook has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.The lumens were able to be flushed prior to placement.The failure mode could not be replicated on the return device.It is possible that bio-matter occluded the lumen, but cook cannot confirm this.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|