• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. VISUALASE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. VISUALASE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 9735559
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received information regarding a thermal therapy system being used during a soft tissue ablation (neuro) procedure.It was reported that there was a tubing leak.There was a leak on the outside portion of the tubing closer to the console.The leak was not close to the catheter.They replaced the console side of the tubing and they were able to proceed with the case.The site had 4 extra tubing sets and opened one of those.There was less than an hour delay to the case.No reported impact to the patient.
 
Additional Manufacturer Narrative
H3,h6) no parts have been received by the manufacturer for evaluation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3/h6: the tube was returned and analyzed.There was a spot on the tubing that appeared to be bent that may have a leak hole.The area has been marked on the tubing for further analysis.Codes b01, c07, and d02 apply.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUALASE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key19019704
Report Number1045254-2024-00471
Device Sequence Number20876150
Product Code GEX
UDI-Device Identifier00763000416423
UDI-Public00763000416423
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K053087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2025
Device Model Number9735559
Device Catalogue Number9735559
Device Lot Number0225978340
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2024
Supplement Date Received by Manufacturer04/04/2024
Initial Report FDA Received Date04/02/2024
Supplement Report FDA Received Date04/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age6 YR
Patient SexMale
Patient Weight22 KG
-
-