| Model Number |
15 |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device had logged event codes that is indicative of a device failure of providing defibrillation energy and associated with monophasic shock.As a result, wrong defibrillation therapy would be delivered, if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was able to duplicate the reported issue.After replacing the therapy pcb assembly, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker product analysis center (pac) evaluated the removed part and determined that the cause of the reported issue was due to a shorted transistor, q8, on the h bridge.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that their device had logged event codes that is indicative of a device failure of providing defibrillation energy and associated with monophasic shock.As a result, wrong defibrillation therapy would be delivered, if needed.There was no patient use associated with the reported event.
|
| |
|