BECTON, DICKINSON & CO., (BD) BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367352
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using bd vacutainer® push button blood collection set with pre-attached holder that there was blood leakage.Personnel was wearing appropriate ppe and none got direct exposure to blood.No patient impact.This occurred 10 times.
 
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for evaluation: yes, h.3 device eval by manufacturer? yes, d9: returned to manufacturer on: 11-apr-2024.Investigation summary: bd received 3 shelf packs of samples for investigation.The samples were evaluated by functional testing and the indicated failure mode for leakage during to injection/blood/urine collection with the incident lot was not observed.Additionally, 30 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to leakage during to injection/blood/urine collection as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode leakage during to injection/blood/urine collection.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that while using bd vacutainer® push button blood collection set with pre-attached holder that there was blood leakage.Personnel was wearing appropriate ppe and none got direct exposure to blood.No patient impact.This occurred 10 times.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key19020728
Report Number1024879-2024-00300
Device Sequence Number7939329
Product Code FPA
UDI-Device Identifier00382903673520
UDI-Public(01)00382903673520
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K220212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Catalogue Number367352
Device Lot Number3292564
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer06/13/2024
Initial Report FDA Received Date04/02/2024
Supplement Report FDA Received Date07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-