| Catalog Number |
367352 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using bd vacutainer® push button blood collection set with pre-attached holder that there was blood leakage.Personnel was wearing appropriate ppe and none got direct exposure to blood.No patient impact.This occurred 10 times.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes, h.3 device eval by manufacturer? yes, d9: returned to manufacturer on: 11-apr-2024.Investigation summary: bd received 3 shelf packs of samples for investigation.The samples were evaluated by functional testing and the indicated failure mode for leakage during to injection/blood/urine collection with the incident lot was not observed.Additionally, 30 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to leakage during to injection/blood/urine collection as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode leakage during to injection/blood/urine collection.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd vacutainer® push button blood collection set with pre-attached holder that there was blood leakage.Personnel was wearing appropriate ppe and none got direct exposure to blood.No patient impact.This occurred 10 times.
|
| |
|
Search Alerts/Recalls
|