COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIA45AMT
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a laparoscopic colon resection, during the fourth firing of functional end-to-end anastomosis, after the left articulation operation was performed while outside the patient cavity, the right articulation started by itself without button operation, and it operated up to about 30 degrees of right articulation.It was discovered that when the right articulation was started, the finger has not been touched with the trigger at all.A new handle and shell were used with the same adapter and cartridge and firing was performed.No patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: sigphandle, sig power sigphandle handle (sn: (b)(6)) sigpshell, sig power sigpshell control shell (lot#n3l1926y) sigadaptstnd, sig power sigadaptstnd linear adapter (sn:unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19020745
Report Number1219930-2024-01403
Device Sequence Number8156656
Product Code GDW
UDI-Device Identifier10884521820555
UDI-Public10884521820555
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEGIA45AMT
Device Catalogue NumberEGIA45AMT
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2024
Initial Report FDA Received Date04/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexUnknown
-
-