Catalog Number 1804250-33
Medical Device Problem Codes Material Separation (1562); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion located in the left anterior descending coronary artery that was 75% stenosed.Reportedly, the 2.5x33mm rx xience skypoint stent delivery system (sds) was advanced; however, failed to cross the lesion due to the anatomy and the proximal shaft separated in two outside the patient.The device was removed without issue.The same size stent was used to complete the procedure.There were no reported adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual and dimensional inspection was performed on the returned device.The reported material separation was confirmed.The reported failure to advance could not be evaluated as the exact anatomical conditions encountered by the device used during the procedure could not be replicated in the test laboratory.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database did not indicate a lot specific quality issue.A conclusive cause for the reported difficulties could not be determined; however, factors that could contribute to failure to advance include, but are not limited to, patient anatomical morphology, patient disease state, pre-dilatation strategy, device placement technique, interaction with previously placed stents or accessory devices.Factors contributing to a material separation include, but are not limited to, damage during manufacturing, interaction with accessory devices or inadvertent mishandling.In this case, it is possible the device may have interacted with the 75% stenosed lesion during advancement, resulting in the reported failure to advance in addition to the observed material deformation and collapsed inner/outer member, ultimately causing the reported material separation; however, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19020827
Report Number2024168-2024-04070
Device Sequence Number3825841
Product Code NIQ
UDI-Device Identifier08717648233111
UDI-Public(01)08717648233111(17)260823(10)3082841
Combination Product (Y/N)Y
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2026
Device Catalogue Number1804250-33
Device Lot Number3082841
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2024
Supplement Date Received by Manufacturer05/14/2024
Initial Report FDA Received Date04/02/2024
Supplement Report FDA Received Date06/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale