Catalog Number EVD35-06-150-120
Medical Device Problem Codes Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Physician attempted to implant an everflex entrust self-expanding stent during treatment of a 150mm calcified lesion in the patient¿s superficial femoral artery (sfa).Artery diameter reported as 6mm.There was no damage noted to packaging (i.E.Shelf carton, hoop/tray).There were no issues noted when removing the device from the hoop/tray.The ifu (instruction for use) was followed during preparation, procedure, post-procedure without issue.Embolic protection was not used.It was reported the stent partially deployed at the target site within the patient vasculature.The thumbscrew/lock-pin was removed prior to attempted deployment.The stent ended up deploying fully by pin-and-pull action after removal of the handle.The stent remains in the patient.There were no attempts made to remove the stent.No deformation was noted to the deployed stent.The delivery system was safely removed.No vessel damage reported.No patient injury reported.
 
Additional Manufacturer Narrative
Product analysis the device was returned with the red safety tab out of the device, there was no stent returned with the device,.The device was confirmed from the strain relief as 150mm, the handle was dismantled, and the pull cable was found to be broken away from the device, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLEX SELF-EXPANDING PERIPHERAL STENT WITH ENTRUST DELIVERY SYSTEM LONG
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19021052
Report Number2183870-2024-00109
Device Sequence Number3879964
Product Code NIP
UDI-Device Identifier00821684051399
UDI-Public00821684051399
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P110023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 06/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/20/2026
Device Catalogue NumberEVD35-06-150-120
Device Lot NumberB628736
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by Manufacturer05/29/2024
Initial Report FDA Received Date04/02/2024
Supplement Report FDA Received Date06/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1644-2022
Patient Sequence Number1
Patient SexUnknown
-
-