| Catalog Number |
366430 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® serum blood collection tubes, an unspecified number of tubes did not clot properly.Sample recollection occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following information has been updated with corrected and/or additional information: d.9.Device available for eval? yes d.9.Returned to manufacturer on: 25-apr-2024 h.6.Investigation summary: bd received seventy-five (75) samples for investigation.The samples were evaluated by visual examination and the indicated failure mode for poor clot formation with the incident lot was not observed.Additionally, eighty (80) retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to poor clot formation as all samples met specifications.Complaints for sample quality are under statistical control for the month of march 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of poor clot formation.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® serum blood collection tubes, an unspecified number of tubes did not clot properly.Sample recollection occurred.
|
| |
|