Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.A2: age at time of event unknown / not provided.A3: gender unknown / not provided.A4: patient weight unknown / not provided.D10: concomitant medical product and therapy dates: oss413, osseotite implant 4 x 13mm, lot number: 2019090905.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|