| Model Number |
32375 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for evaluation.It was observed that the suture returned inside the stent.It was possible to observe that the suture was detached.No other damages were observed.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 06mar2024.It was reported that the device was difficult to advance.An 8f flexima nephrostomy catheter system was selected for use.During the procedure, a hard resistance was encountered upon pushing the catheter in.The procedure was completed with a different device.No complications reported, and the patient's condition was normal.However, investigation of the returned device revealed that the suture was detached.
|
| |
|