Model Number 602
Medical Device Problem Codes Off-Label Use (1494); Use of Device Problem (1670)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 03/14/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Lens noted by customer to be scratched at completion of procedure.Lens tore in half and fell to back of eye when attempting to explant the lens.Patient was left aphakic.
 
Additional Manufacturer Narrative
The customer stated that the device cannot be returned therefore a proper root cause analysis could not be completed nor the reported issue confirmed.Without a proper device analysis, a definitive root cause cannot be determined at this time.It was stated by the customer that they are using the yamane technique to implant the lenses.This technique induces a larger amount of force while trying to position the haptics in the scleral tunnel.Our lenses are intended to be implanted in the capsular bag.Thus, the lens was being used off-label.There was no damage reported to have been noticed during the inspection prior to use and preparation, which suggests a product deficiency, did not contribute to the reported difficulties.Based on the information provided, the risk assessment for the lucia product, and our experience we have determined that the following factors may have cause or contributed to the reported issue but not limited to: lens placement technique, loading strategy, and poor handling during folding and inserting.The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lens that would have caused or contributed to the nature of this complaint.Additionally, our lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operating procedures (b)(4) and inspections and have met all criteria for release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CT LUCIA
Common Device Name
CT LUCIA 602
Manufacturer (Section D)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s. vintage ave. bld. a
ontario CA 91761
Manufacturer (Section G)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s. vintage ave. bld. a
ontario CA 91761
Manufacturer Contact
manjaya hegde
5300 central parkway
dublin, CA 94568
9252164697
MDR Report Key19021816
Report Number3010126268-2024-00042
Device Sequence Number11714750
Product Code HQL
UDI-Device Identifier00843045100980
UDI-Public(01)00843045100980(17)280229
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
P10016-S008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number602
Device Catalogue Number003500-0050-677
Device Lot Number3S230281
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2024
Initial Report FDA Received Date04/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-