Catalog Number 1000184
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a lesion in the mid right coronary artery (mrca) with moderate calcification and moderate tortuosity.During pre-dilatation, the indeflator was noted to have a leak in the pressure gauge housing.Another 20/30 indeflator was used to continue the procedure with the same dilatation balloon.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As the device was not returned for analysis the investigation was unable to determine a conclusive cause for the reported leak difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20/30 PRIORITY PACK INDEFLATOR
Common Device Name
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19022554
Report Number2024168-2024-04079
Device Sequence Number3880506
Product Code MAV
UDI-Device Identifier08717648013591
UDI-Public(01)08717648013591(17)241130(10)60418865
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K961471
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Catalogue Number1000184
Device Lot Number60418865
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/11/2024
Supplement Date Received by Manufacturer04/02/2024
Initial Report FDA Received Date04/02/2024
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.5X15MM EUPHORA BALLOON; SION GUIDE WIRE
Patient SexUnknown
-
-