| Catalog Number |
05P58-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4) apoc labeling was evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 27-mar-2024, abbott point of care (apoc) was contacted by a customer regarding i-stat b-hcg cartridges that yielded a positive result on a 13 year old female patient with osteogenesis imperfecta.There was no additional patient information at the time of this report.Return product is available for investigation.Method date tested results sample i-stat (b)(6) 2024 09:26 30.3 1 i-stat (b)(6) 2024 09:50 23.7 2 there are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.I-stat positive cut-off values: b-hcg < 5.0 iu/l negative 5.0 < ss-hcg < 25.0 iu/l indeterminate b-hcg >25.0 iu/l positive intended use: the i-stat® total beta-human chorionic gonadotropin (b-hcg) test is an in vitro diagnostic test for the quantitative and qualitative determination of b-hcg in venous whole blood or plasma samples using the i-stat 1 analyzer systems.The test is intended to be used as an aid in the early detection of pregnancy and is for prescription use only.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 29-apr-2024.A review of the device history record confirmed the cartridge lot met finished goods release criteria.Cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.An (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for total b-hcg cartridge lot f23311a.
|
| |
|
Search Alerts/Recalls
|