| Catalog Number |
1012271-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed de novo lesion in the distal left anterior descending artery (dlad) with moderate calcification and moderate tortuosity.The 2.25x15mm trek balloon dilatation catheter (bdc) was advanced for pre-dilatation; however, the balloon ruptured at 12 atmospheres (atm) during the first inflation.The bdc was removed from the patient.A non-abbott bdc was used to continue the procedure.There was no adverse patient effect and there was no reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The electronic lot history record and exception reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.The investigation determined that the reported balloon rupture appears to be related to operational context.Based on the reported information, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|