• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804300-08
Medical Device Problem Codes Stretched (1601); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.Attachment: medwatch report #mw5152005.
 
Event or Problem Description
User facility medwatch report [mw5152005] received that states: catheter defective at junction of rx.It was reported prior to the procedure, the 3.0x08mm xience skypoint stent was noted to be defective and started to unravel.The device was not used.An unspecified device was used to complete the procedure.There was no patient involvement.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database did not indicate a lot specific quality issue.A conclusive cause for the reported difficulties could not be determined; however, factors that could contribute to a stretched stent include, but are not limited to, damage during manufacturing, inadvertent mishandling during sheath/stylet removal, preparation, during use of the device, interaction with accessory devices, patient anatomical morphology, or patient disease state.Material deformation may be attributed to several factors including, but are not limited to, damage during manufacturing, inadvertent mishandling during unpackaging, sheath/stylet removal, during preparation or during use of the device.Stent dislodgement may be attributed to several factors including, but not limited to, improper or inadequate crimping at the time of manufacture, incorrect sheath sizing, positive pressure during preparation, forced sheath removal, handling of the stent during preparation, and interaction with accessory devices.Based on the information provided and without the device to examine, a definitive cause for the reported stretched stent, material deformation and stent dislodgement cannot be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19026965
Report Number2024168-2024-04089
Device Sequence Number8483382
Product Code NIQ
UDI-Device Identifier08717648233210
UDI-Public(01)08717648233210(17)241116(10)111094A
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2024
Device Catalogue Number1804300-08
Device Lot Number111094A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date04/02/2024
Supplement Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-