• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VENOUS WALLSTENT; STENT, ILIAC VEIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VENOUS WALLSTENT; STENT, ILIAC VEIN Back to Search Results
Model Number H74912044166070
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Chest Pain (1776); Valvular Insufficiency/ Regurgitation (4449)
Date of Event 10/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3: date of event estimated using date article was received for publication.Khalid mo, moskovits n, frankel ra, et al.Venous stent migrating to the right heart causing severe regurgitation.Journal of investigative medicine high impact case reports.2020;8.Doi:10.1177/2324709620974220.
 
Event or Problem Description
It was reported via literature article that stent migration occurred.A 16x60mm venous wallstent was implanted in the left iliac vein 10 months prior to presentation.Two days after the procedure, the patient went to a hospital complaining of chest pain.Transesophageal echocardiography was performed at the other institution and showed an "unclear echo density in the right atrium, which was suspected to be an artifact." a coronary angiogram showed nonobstructive coronary artery disease.Based on the previous report, a transthoracic echocardiography was ordered that showed a foreign body in the right atrial cavity extending across the tricuspid valve (tv).A transesophageal echocardiography was performed, which showed a dislodged vascular stent in the right atrium measuring 5.5 x 1.6 cm.The stent was impinging on the tv, causing severe tricuspid regurgitation.The patient was admitted to cardiology service and placed on heparin for anticoagulation.Computed tomography of the chest reported a tubular radiopaque density, in the right atrium, partially extending into the right ventricle.Following a multidisciplinary meeting involving interventional cardiology, vascular surgery, and cardiothoracic surgery, percutaneous extraction of the stent was attempted under transesophageal echocardiography guidance; however, it was firmly adhered to the tv and opposing the right atrial wall.Consequently, the patient underwent open stent extraction, during which the stent was found to be adherent to the tv and the right atrial wall.The stent was extracted via open thoracotomy.The extracted stent measured 6cm in length with white tissue observed on it, which was noted to be part of the tv endothelium.The patient required tv replacement and was discharged in a stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOUS WALLSTENT
Common Device Name
STENT, ILIAC VEIN
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19028642
Report Number2124215-2024-18833
Device Sequence Number8187302
Product Code QAN
UDI-Device Identifier08714729979579
UDI-Public08714729979579
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980033/S050
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74912044166070
Device Catalogue NumberH74912044166070
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date04/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age49 YR
Patient SexMale
-
-