ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 14" (36 CM) APPX 3.6 ML, TRANSFER SET W/4 CLAVE¿, 2 TRI-CONNECTORS, BACK CHECK V; STOPCOCK, I.V. SET
|
Back to Search Results |
|
| Catalog Number |
B33973 |
| Medical Device Problem Codes |
Disconnection (1171); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available for investigation as the customer has discarded it.Without the return of the device, a probable cause is unable to be determined.
|
| |
|
Event or Problem Description
|
|
The event involved a 14" (36 cm) appx 3.6 ml, transfer set w/4 clave¿, 2 tri-connectors, back check valve, clamp (blue), vented cap where the customer reported that they had a defective tubing set that disconnects or separates.There was patient involvement but harm was not reported as a consequence of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product samples, pictures, or videos were received for investigation, however, the complaint of a separation on the b33973 can be confirmed based on correction actions that are current in place for product.The probable cause is due to a manual bonding error in manufacturing.Lot history review for lot 5916735 was reviewed and no relevant non conformities were found that would have led to the reported complaint.
|
| |
|