| Catalog Number |
B33974 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Chemical Exposure (2570); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a 21" (53 cm) appx 5.4 ml, transfer set w/8 clave¿, 4 tri-connectors, back check valve, clamp (blue), vented cap where it was reproted the 8-port manifold was infusing mycophenylate when another manifold port with used tubing for ganciclovir disconnected below needleless connector.This caused hazardous medication spill of mycophenylate ~100 ml.Spill cleaned with spill kit and surface safe per policy.They did not sequester per material services request as infusing hazardous meds, thus no sample is returning for testing.There was patient involvement and unknown human harm however no other details were provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available for investigation.Without the return of the device a probable cause is unable to be determined.Possible lot numbers: 13833492 manufactured date 12/2023.1386766 manufactured date 01/2024.
|
| |
|
Additional Manufacturer Narrative
|
|
A photo was returned showing the clave missing from the bond pocket of the tri-connector.The reported complaint of separation can be confirmed on the b33973 transfer set w/4 clave.The probable cause is due to a manual bonding error.
|
| |
|
Event or Problem Description
|
|
Additional information provided on 20may2024 via medwatch (b)(4) there was also precipitation observed down the iv tubing, which was caught before it reached patient.It is unclear whether this was due to the manifold defect however, mycophenolate and ganciclovir are not compatible.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: related report number added to h10 versus in b5 from previous supplemental report.
|
| |
|