• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 XTND GLENO D42MM +2MM; SHOULDER IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 XTND GLENO D42MM +2MM; SHOULDER IMPLANT Back to Search Results
Catalog Number 130762142
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the surgeon opened standard metaglene in a reverse shoulder and implanted it.Then opened a glenosphere to attach to meteglene.The glenosphere would not advance as if something was stopping threads.A new baseplate metaglene was opened and still no advancement would happen.A new glenophere was opened and with new metaglene and glenosphere the implants were successfully seated.It is thought that threads in first implanted metaglene were damaged and that is why glenoshere would not seat.This then damaged the glenosphere threads surgery delayed 15 minutes due to the reported event.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: surgeon opened standard metaglene in a reverse shoulder and implanted it.Then opened a glenosphere to attach to meteglene.The glenosphere would not advance as if something was stopping threads.A new baseplate metaglene was opened and still no advancement would happen.A new glenophere was opened and with new metaglene and glenosphere the implants were successfully seated.It is thought that threads in first implanted metaglene were damaged and that is why glenoshere would not seat.This then damaged the glenosphere threads the device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned device found the leading screw was damaged from the threaded area, which is indicative of being cross threaded.This type of damage would explain why the surgeon could not properly mate the components.Additionally, the device presents signs of what appears to be organic material at the bottom surface.Please refer to the surgical technique delta xtend¿ reverse shoulder system, page 34: proper alignment between the glenosphere and metaglene is absolutely essential to avoid cross threading between the components.If the glenosphere seems difficult to thread onto the metaglene, do not force engagement but re-align the components.If necessary, remove the inferior retractor or improve the capsular release.It is important to check that there is no soft tissue between the metaglene and glenosphere.A dimensional inspection for the xtnd gleno d42mm was unable to be performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the xtnd gleno d42mm +2mm would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation were performed for the finished device 130762142, lot number d22032024, it was manufactured on 23-mar-2022.(b)(4) parts were manufactured per specification and all raw materials met specification, and no non-conformances / manufacturing irregularities were identified.Device history review : a manufacturing record evaluation were performed for the finished device 130762142, lot number d22032024, it was manufactured on 23-mar-2022.(b)(4) parts were manufactured per specification and all raw materials met specification, and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTND GLENO D42MM +2MM
Common Device Name
SHOULDER IMPLANT
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19030794
Report Number1818910-2024-07326
Device Sequence Number12771506
Product Code PHX
UDI-Device Identifier10603295502500
UDI-Public(01)10603295502500(17)270228(10)D22032024
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number130762142
Device Lot NumberD22032024
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2024
Supplement Date Received by Manufacturer04/24/2024
05/08/2024
05/09/2024
07/22/2024
Initial Report FDA Received Date04/03/2024
Supplement Report FDA Received Date04/25/2024
05/09/2024
05/15/2024
07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND METAGLENE
Patient SexUnknown
-
-