• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US GRYPHON P BR ANCHOR W/OC; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US GRYPHON P BR ANCHOR W/OC; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 210814
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the healthcare professional from china that during a labral repair procedure, it was discovered that the gryphon p br anchor w/oc device broke off.All the broken parts were removed.It was not reported if there were any delays to the surgical procedure.A spare device was used to complete the surgery.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: h4: the device manufacture date is currently unavailable.E1: the reporter¿s complete facility address was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: h4: the device manufacture date was unavailable in the initial medwatch report.The device manufacture date had been identified and updated accordingly.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device lot number (166l214), and no non-conformance was identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRYPHON P BR ANCHOR W/OC
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19031045
Report Number1221934-2024-01053
Device Sequence Number18151583
Product Code MAI
UDI-Device Identifier10886705001286
UDI-Public10886705001286
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2026
Device Catalogue Number210814
Device Lot Number166L214
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2024
Supplement Date Received by Manufacturer03/27/2024
Initial Report FDA Received Date04/03/2024
Supplement Report FDA Received Date04/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-