|
DEPUY SYNTHES PRODUCTS LLC KINCISE¿ ANTERIOR BROACH ADAPTER; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
| Catalog Number |
1010-01-102 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: d10: concomitant med products and therapy dates: actis broach, device, (b)(6) 2024.E1: the reporter¿s phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the kinsise adapter device got stuck on the broach device and could not be removed.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H6: device history review: the actual device was not returned for evaluation; therefore, the reported condition was not confirmed.However, a device history review was performed which indicated that there were no non-conformances or abnormalities identified during the manufacture of the device that may have contributed to the reported condition.
|
| |
|
Search Alerts/Recalls
|
|
|