• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. XIENCE SKYPOINT¿; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. XIENCE SKYPOINT¿; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 1804225-12
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Angina (1710); Non specific EKG/ECG Changes (1817); Thrombosis/Thrombus (4440)
Date of Event 03/19/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified, moderately tortuous first diagonal branch.The 2.25x12mm xience skypoint stent was implanted without issue; however, the patient began to experience chest pain, which was confirmed with positive ekg.Angiography was performed and observed acute thrombus had occurred.The stent was ballooned and thrombectomy was performed on the diagonal lesion.It was noted that the correct stent size was selected; however, may have been not been fully apposed to vessel.There was no adverse patient sequela and no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no similar complaints from this lot.A conclusive cause for the reported patient-device incompatibility (wall apposition) could not be determined.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the subsequent unexpected medical intervention appears to be related to the operational context of the procedure.Patient-device incompatibility (wall apposition) may be attributed to several factors including, but not limited to, patient anatomical morphology, deployment technique, product size selection, or interference from previously deployed devices.In this case, it is possible that the stent may have been undersized for the vessel, as reported by the physician; however, this cannot be confirmed.The reported patient effects of angina and thrombosis are listed in the xience skypoint, everolimus eluting coronary stent system (eecss), electronic instructions for use as known patient effects of coronary stenting procedures.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT¿
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19033448
Report Number2024168-2024-04162
Device Sequence Number15214327
Product Code NIQ
UDI-Device Identifier08717648232985
UDI-Public(01)08717648232985(17)240613(10)106034A
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/13/2024
Device Model Number1804225-12
Device Catalogue Number1804225-12
Device Lot Number106034A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/19/2024
Supplement Date Received by Manufacturer08/20/2024
Initial Report FDA Received Date04/03/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexMale
-
-