Catalog Number A2025-060
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6: medical device problem code 2017 clarifier- incorrect prep.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified, mildly tortuous peroneal artery that is 80% stenosed.The 2.5x60mm armada 14 balloon catheter was not air aspirated outside the anatomy prior to use.Once at the lesion, the balloon would not hold pressure after inflation to nominal pressure and was noted to be deflating.However, balloon angioplasty was completed with the same armada and then the device was removed.There was no adverse patient effects and no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported that the device was prepped inside the anatomy.It should be noted that the percutaneous transluminal angioplasty (pta), armada 14, global instructions for use (ifu) specifies: to perform the steps outlined 1 through 7 to remove all air and verify the integrity of the pta catheter.The instructions indicate to perform the steps outside of the patient¿s anatomy.In this case, it is unknown if the ifu violation caused or contributed to the reported complaint.The investigation determined that the reported balloon rupture appears to be related to operational context.In this case, it is likely that the balloon interacted with the moderately calcified and mildly tortuous anatomy such that the balloon became damaged and subsequently ruptured during inflation.Based on the reported information, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.D9: device available for evaluation updated from ¿yes¿ to ¿no¿ h3: device evaluated by mfg? updated from ¿yes¿ to ¿no¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19033495
Report Number2024168-2024-04164
Device Sequence Number15208684
Product Code DQY
UDI-Device Identifier08717648156021
UDI-Public(01)08717648156021(17)260630(10)3012341
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue NumberA2025-060
Device Lot Number3012341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/19/2024
Supplement Date Received by Manufacturer06/18/2024
Initial Report FDA Received Date04/03/2024
Supplement Report FDA Received Date06/21/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-