|
OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 6,4/7,8 FR. X 330 MM, 7°, 4,2 FR. CHANNEL, WITH WA00395A
|
Back to Search Results |
|
| Model Number |
WA29041A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the device has not been returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the ureteroscope, 6,4/7,8 fr.X 330 mm, 7°, 4,2 fr.Channel, with wa00395a exhibited a broken component.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to considerable mechanical force caused by an impact or fall.There are no countermeasures necessary for the suggested phenomenon because the associated risk is acceptable.The manufacturer will monitor the occurrence rate in the context of the utilized quality management system.If necessary, the manufacturer will take action in the future.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|