| Model Number |
BEA28-90/I20-30 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Clinical Endoleak (4843)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2 h3 other text : device remains implanted.
|
| |
|
Event or Problem Description
|
|
The patient underwent an endovascular aneurysm repair (evar) on (b)(6) 2022 with the implant of an afx2 bifurcated stent graft, an afx vela suprarenal, two (2)excluder legs, and a liminexx stent.This initial procedure is outside of the indications of use (off-label) due to the use of adjunctive devices not compatible with the afx2 system.Approximately one year post initial procedure the aneurysm had grown from 51mm to 56mm.Approximately 1.5 years post initial procedure, during a routine follow-up with computed tomography (ct) and intraoperative angiography, the aneurysm measured 62mm which the physician suspects is due to a type iiib endoleak.Reintervention is scheduled to be on (b)(6) 2024.
|
| |
|
Event or Problem Description
|
|
The patient underwent an endovascular aneurysm repair (evar) on (b)(6) 2022 with the implant of an afx2 bifurcated stent graft, an afx vela suprarenal, two (2)excluder legs, and a liminexx stent.This initial procedure is outside of the indications of use (off-label) due to the use of adjunctive devices not compatible with the afx2 system.Approximately one year post initial procedure the aneurysm had grown from 51mm to 56mm.Approximately 1.5 years post initial procedure, during a routine follow-up with computed tomography (ct) identified that the aneurysm measured 62mm and a possible type iiib endoleak.The re-operation was performed on (b)(6) 2024.On angiography aneurysm was show to continue to grow and endoleak was not confirmed.The physician elected to reline afx2 bifurcated stent graft with or excluder conformable system.Patient status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix the aneurysm enlargement, possible type iiib endoleak and additional endovascular procedure complaints are unconfirmed.Device, user, procedure or anatomy relatedness of this complaint could not be determined.No procedure related harms for this complaint were identified.There was concomitant product use, it is unclear if this contributed to the reported event.The final patient status was reported to be post additional endovascular procedure done (b)(6) 2024.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2 corrections: g3- awareness date: updated h6- investigation finding codes: removed 3233 h6- investigation conclusion codes: removed 11.
|
| |
|