Model Number 2532
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 evaluation of the returned product noted 21 pairs were received from 8 different lot numbers.There are 4 different lots (15 pairs) contained male buckles with slipping issue if overhand knot is not present.When the application instructions prescribed in the ifu were followed, the restraints did not exhibit slippage issue indicating the product is functional and will perform as intended.However, if the ifu step 1b.Was not followed where the overhand knot with the excess strap was not present to limit unwanted adjustment, the bed connecting strap exhibited slippage due to the male buckle.Possible root cause for this deficiency is that the ifu was not followed while applying this restraint.There are 4 lots (6 pairs) contained male buckles with no slipping issue even if the overhand knot is not present.A review of the complaint database revealed similar events against this device in the past two years leading tidi to redesign the male buckle of the product to mitigate the slipping issue from occurring while the knot is not present.The corrective action is being addressed via (b)(4).The ifu application instructions state to attach the female end of the quick-release buckle (short strap) to the frame that moves with the patient, out of the patient¿s reach (do not attach to side rail or head/footboard).You may also wrap the connecting strap once around the frame to move the buckle out of the patient¿s reach.Secure by feeding the female end through the loop in the strap.Insert the male end of the connecting strap into the female end of the short strap.Listen for a snapping sound, pull firmly on straps to ensure a good connection.To limit unwanted adjustment, tie an overhand knot with the excess strap directly below the quick-release buckle.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
 
Event or Problem Description
Customer reporting a complaint on product # 2532.Customer reported buckle issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOAM LIMB HOLDER
Common Device Name
RESTRAINT, PROTECTIVE
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key19035376
Report Number2182318-2024-00041
Device Sequence Number6986065
Product Code FMQ
UDI-Device Identifier10190676004789
UDI-Public10190676004789
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2532
Device Catalogue Number2532
Device Lot Number3339T019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2024
Initial Report FDA Received Date04/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-