Model Number PMG-ADVANCED
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code Burn(s) (1757)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The actual device is reported to be available but has not been returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 03-apr-2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint: (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.H3 other text: device not returned.
 
Event or Problem Description
It was reported the "machine was not able to achieve burning temperature, with impedance rising from 500 to 2200 ohms prior to self-terminating after approximately 40 sec attempted further 2 attempts at rf at above settings with same issue pulled back blanket to ensure diathermy pad was well adhered to the leg in case that was the issue and noted charring.Immediately turned off the machine and instructed [the physician] to terminate the procedure." the patient's injuries noted two points of burn on the corresponding to diagonal points of the diathermy pad - two burns on thigh with charring."unclear if arced over skin or through muscle.Medical treatment was provided secondary to burns to thighs.The clinicians "impression is that there was poor contact of the skin adhesive to the skin due to interference from hair." medical treatment "was provided secondary to burns to thighs.".
 
Additional Manufacturer Narrative
Correction: d1 all information reasonably known as of 07-jun-2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, st8003-008, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of (b)(6) 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PMG-ADVANCED V4.2A
Common Device Name
PAIN MANAGEMENT GENERATORS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
STELLARTECH RESEARCH CORP.
560 cottonwood dr
milpitas CA 95035
Manufacturer Contact
ujjal chakravartty
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key19037430
Report Number3011270181-2024-00043
Device Sequence Number7338987
Product Code GEI
UDI-Device Identifier10680651422979
UDI-Public(01)10680651422979(20)02(21)ST8003-008
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K163461
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPMG-ADVANCED
Device Catalogue NumberN/A
Device Lot NumberST8003-008
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by Manufacturer05/09/2024
07/23/2024
Initial Report FDA Received Date04/03/2024
Supplement Report FDA Received Date06/08/2024
08/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/17/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexMale
Patient Weight71 KG
-
-