| Catalog Number |
MS18501 |
| Medical Device Problem Codes |
No Device Output (1435); No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that there was no alarm at the ics when an ecg or spo2 cable is disconnected from the m300.The user at the central will potentially will not know that the patient is no longer being monitored which has the potential for adverse patient impact.No injury or other patient impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported to draeger that when the m300 cable disconnected there was no alarm.The available ics logs and a movie of the event occurring were analyzed by draeger engineering.The movie and the ics logs do confirm alarming for the cited bed (tele 1) but no alarms were generated during the reported date of event (3-15-2024 @ 15:21) for this m300 device.The m300 alarm logs are necessary to further investigate the cause of the issue.Unfortunately the logs were unavailable for this investigation.Based on the evidence provided to investigate this issue, a root cause could not be determined for the reported behavior.
|
| |
|
Event or Problem Description
|
|
The customer reported that there was no alarm at the ics when an ecg or spo2 cable is disconnected from the m300.The user at the central will potentially will not know that the patient is no longer being monitored which has the potential for adverse patient impact.No injury or other patient impact was reported.
|
| |
|