| Model Number |
328438 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Material # : 328438.Material description: syringe 0.3ml 31g 8mm wholeunit 10bg 500.Complaint description: customer has been having issues with these syringes for quite some time.Customer tossed a lot out, thinking it was just a bad batch but now keeps on happening.The bolder writing syringes are harder to push out once filled with toxins.Provider note: difficult to push out smaller amounts of medication extremely difficult to control push.Wrong dosages due to issue.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 other text : device is not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no physical samples were received however the investigation was performed based on the photo(s) provided.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Embecta was unable to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situation analysis (sa) is required at this time.
|
| |
|
Search Alerts/Recalls
|