MAUDE Adverse Event Report: STRYKER GMBH LONG NAIL KIT R1.5, TI, LEFT GAMMA3 Ø13X380MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LONG NAIL KIT R1.5, TI, LEFT GAMMA3 Ø13X380MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 35253380S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.H3 other text: device disposition is unknown.
 
Event or Problem Description
As reported: "gamma 3 implanted 16 months ago and 100 days later the implant broke.Screws at the bottom broke and nail also broke by the point where the lag screw is implanted.Entire construct was revised and patient experienced a similar event 100 days later.".
 
Additional Manufacturer Narrative
The reported event could be confirmed based on details provided, only.Images available revealed a broken nail.Due to missing device a physical evaluation was impossible and further technical statement could not be given.In the case presented a 61-year-old male patient, at time of reported event, was primarily treated with a long nail kit r1.5, ti, left gamma3 013x380mm x 125 on (b)(6) 2022 and revised on (b)(6) 2023 by a long nail kit r1.5, ti, left gamma3 015x400mm x 125 during a 1st revision.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
 
Event or Problem Description
As reported: "gamma 3 implanted 16 months ago and 100 days later the implant broke.Screws at the bottom broke and nail also broke by the point where the lag screw is implanted.Entire construct was revised and patient experienced a similar event 100 days later.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONG NAIL KIT R1.5, TI, LEFT GAMMA3 Ø13X380MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19038219
Report Number0009610622-2024-00126
Device Sequence Number6992361
Product Code HSB
UDI-Device Identifier04546540409928
UDI-Public04546540409928
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2025
Device Catalogue Number35253380S
Device Lot NumberK0DD2F3
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer05/31/2024
Initial Report FDA Received Date04/03/2024
Supplement Report FDA Received Date06/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight77 KG
-
-