| Model Number |
EQUINOXE CAGE GLENOID SMALL, ALPHA |
| Medical Device Problem Codes |
Fracture (1260); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the 64 year old female patient had an initial right tsa on (b)(6) 2020.The patient was revised on (b)(6) 2024 due to poly failure.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.Photos & x-rays received.The devices are not available for evaluation due to chain of command.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products: - equinoxe, humeral stem primary, press fit 9mm (cat# 300-01-09 / serial# (b)(6)).- equinoxe replicator plate 4.5mm o/s (cat# 00-10-45 / serial# (b)(6)).- equinox square torque define screw drive kit (cat# 300-20-02 / serial# (b)(6)).- equinoxe, humeral head short, 38mm (alpha) (cat# 310-01-38 / serial# (b)(6)).- equinoxe cage glenoid small, alpha (cat# 314-13-02 / serial# (b)(6)).- glnd kwire (cat# 315-35-00 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
The revision was likely because of the glenoid component migrating from its intended location due to cage fracture and loosening.However, the potential cause of the failure and the sequence of events cannot be confirmed as the device was not returned for evaluation.This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6 clinical code and problem code.
|
| |
|