| Model Number |
ACL TIGHTROPE RT W/ DEPLOYING SUTURE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 3/25/2022 it was reported by a sales representative via sems that an ar-1588rt-j acl tightrope had an issue when implanting.The graft went up in the femoral tunnel but while trying to tension the mechanism, the suture broke and, therefore; the implant was unusable.
|
| |
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The probable cause of the event could not be determined from the information available and without device evaluation.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used when tensioning the sutures.
|
| |
|