| Catalog Number |
07026935190 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) performed repeatability testing for anti-tpo and confirmed there were no issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Calibration and qc were acceptable.No issues were identified during a review of the alarm trace data.Due to the limited information provided, the cause of the event could not be determined.The investigation did not identify a product problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The e801 module serial number is (b)(6).
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of a discrepant high result for 1 patient sample tested for elecsys anti-tpo (anti-tpo) on a cobas 8000 e 801 module.The initial result was 73.2 iu/ml.The sample was repeated twice with results of less than 9 iu/ml.The repeat results were believed to be correct as these matched the patient¿s previous anti-tpo results.
|
| |
|