Catalog Number 07026935190
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service engineer (fse) performed repeatability testing for anti-tpo and confirmed there were no issues.
 
Additional Manufacturer Narrative
Calibration and qc were acceptable.No issues were identified during a review of the alarm trace data.Due to the limited information provided, the cause of the event could not be determined.The investigation did not identify a product problem.
 
Additional Manufacturer Narrative
The e801 module serial number is (b)(6).
 
Event or Problem Description
The initial reporter complained of a discrepant high result for 1 patient sample tested for elecsys anti-tpo (anti-tpo) on a cobas 8000 e 801 module.The initial result was 73.2 iu/ml.The sample was repeated twice with results of less than 9 iu/ml.The repeat results were believed to be correct as these matched the patient¿s previous anti-tpo results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-TPO
Common Device Name
SYSTEM, TEST, THYROID AUTOANTIBODY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19041294
Report Number1823260-2024-01035
Device Sequence Number7270898
Product Code JZO
UDI-Device Identifier04015630939626
UDI-Public04015630939626
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K051890
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Catalogue Number07026935190
Device Lot Number750634
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/13/2024
Supplement Date Received by Manufacturer04/10/2024
04/19/2024
Initial Report FDA Received Date04/04/2024
Supplement Report FDA Received Date04/12/2024
04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-