BOSTON SCIENTIFIC CORPORATION INTELLATIP MIFI XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 87165
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Headache (1880)
Date of Event 03/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that the catheter tip detached from the rest of the catheter and was unable to be retrieved, which required a surgical intervention.Per literature review, it was reported that: a 52-year-old man treated with previous mechanical mitral valve replacement presented to an outside hospital for planned cardiac ablation for atrial fibrillation (af) using an intellatip mifi xp ablation catheter.During the procedure, while advancing the intellatip ablation catheter across the mechanical mitral valve, the catheter was sheared, leaving the distal tip intravascularly and unable to be retrieved.He was emergently transferred to the advocate health care facility, united states, for further workup and treatment.On arrival, workup including chest x-ray, abdominal x-ray and neck x-ray failed to reveal the retained catheter tip.Therefore, a skull x-ray was performed and then confirmed with head computed tomography (ct) showing the catheter tip present in the right m1 segment of the middle cerebral artery.The patient arrived neurologically intact with only symptoms of a mild headache.Due to concerns of developing a large vessel occlusion, patient was emergently taken to the neurointerventional suite for planned removal of the retained catheter tip.Catheter tip was successfully retrieved during neurosurgery.Follow-up angiogram showed no further retained catheter without occlusion or stagnation of flow of any of the middle cerebral artery branches.The retained cardiac catheter tip was retrieved approximately 12 hours after the initial planned cardiac ablation procedure.The patient made a complete recovery without any neurological deficits or evidence of infarct on follow-up imaging.Young, m., johnson, r., & farhat, h.(2020).Monorail microsnare retrieval of cardiac microcatheter tip embolised to the middle cerebral artery after attempted cardiac ablation procedure.Bmj case reports, 13(3).Https://doi.Org/10.1136/bcr-2020-015800.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI XP
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key19041723
Report Number2124215-2024-19667
Device Sequence Number7222868
Product Code OAD
UDI-Device Identifier08714729848011
UDI-Public08714729848011
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P020025
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Literature
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number87165
Device Catalogue Number87165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/11/2024
Initial Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexMale
-
-