• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH NCB®, PERIPROSTHETIC FEMUR PLATE, DISTAL, LEFT, 18 HOLES, 355 MM; TRAUMA, PLATE.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH NCB®, PERIPROSTHETIC FEMUR PLATE, DISTAL, LEFT, 18 HOLES, 355 MM; TRAUMA, PLATE. Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: report source australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient underwent revision approximately 6 months after implantation due to a fracture of the implanted plate.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned for evaluation, however two pictures was provided.One picture shows the explanted plate, screws, and cerclage wires.The other picture is showing the reference number of the plate.As the plate appears to be fractured, event can be confirmed.A review of the device manufacturing records could not be performed due to missing lot number.Review of the complaint history identified additional similar complaints for the reported item.However, due to the lack of the lot number, an additional lot search could not be performed.Complaints are monitored per complaint trending process in order to identify potential adverse trends.Two ap views of the left femur (the provided x-ray pictures were reviewed from a radiologist): left hip and knee arthroplasties are anatomically aligned.There is extensive lateral plate, multilevel screw, and cerclage wire fixation of the femur.The lateral plate is fractured at the mid-femur level with slight lateral angulation and the femur is fractured.Bone quality appears markedly osteopenic.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCB®, PERIPROSTHETIC FEMUR PLATE, DISTAL, LEFT, 18 HOLES, 355 MM
Common Device Name
TRAUMA, PLATE.
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key19042350
Report Number0009613350-2024-00128
Device Sequence Number12879568
Product Code HRS
UDI-Device Identifier00889024292918
UDI-Public(01)00889024292918
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K100111
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number0202264118
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2024
Supplement Date Received by Manufacturer06/11/2024
Initial Report FDA Received Date04/04/2024
Supplement Report FDA Received Date06/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-