| Catalog Number |
011-CS25 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available for investigation.Photos were provided by the customer.Investigation is pending.Initial reporter address: (b)(6).
|
| |
|
Event or Problem Description
|
|
The event involved a check valve with male/female luer lock, where it was reported in the infusion line, blood returns through the non-return valve.The event occurred in the anesthesia operating room when the product was placed on the infusion line.The valve was changed.The customer stated there was non clinically significant blood loss.There was a delay in therapy due to the time to change valve and bandage.The therapy was completed.There was no drug administered, but contrast agent.The doctor had some blood on his hands.There was no medical intervention needed, just hand washing.The leak did not come into contact with the patient.The leak was cleaned up according to facility protocol with no specific kit.The patient¿s condition before, during and after the incident was under general anesthesia.The patient received the full intended dose after changing the consumable.There was patient involvement, however, no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Received one new 011-cs25, check valve with male/female luer lock for inspection.No defects or anomalies noted.The set was primed at gravity pressure per packaging directions.The back pressure was tested per product specifications.There was no back flow observed.The reported complaint was unable to be replicated or confirmed.The lot history was reviewed, and no nonconformities were identified that may have contributed to the reported complaint.D9 device returned to manufacturer on 4/22/2024.
|
| |
|
Search Alerts/Recalls
|