• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ORGANOX LIMITED ORGANOX METRA; NORMOTHERMIC MACHINE PERFUSION SYSTEM FOR THE PRESERVATION OF DONOR LIVERS PRIOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ORGANOX LIMITED ORGANOX METRA; NORMOTHERMIC MACHINE PERFUSION SYSTEM FOR THE PRESERVATION OF DONOR LIVERS PRIOR Back to Search Results
Model Number D0146
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 03/05/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The liver was recovered, cannulated, and connected to the device by a surgeon at the organ procurement organization (opo).During a follow up phone call after the transplant, a member of the clinical team spoke with a surgeon at the implanting transplant facility.The surgeon believed that the arterial cannula had been inserted too far into the hepatic artery, at approximately the depth of the gastroduodenal artery.On removal of the arterial cannula, a thrombus was seen in the hepatic artery.The implanting surgeon was able to trim the vessel and continue with the transplant and hepatic arterial anastomosis.The initial reporter was unable to provide the device lot number.The arterial cannula is believed to have been discarded by the hospital that transplanted the donor liver, thus it could not be evaluated.The root cause of the reported issue is undetermined.
 
Event or Problem Description
A surgeon reported that "upon removal of the arterial cannula, there was a thrombosis.The liver was still transplanted, but it did increase the time on ice by approximately 30 minutes".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORGANOX METRA
Common Device Name
NORMOTHERMIC MACHINE PERFUSION SYSTEM FOR THE PRESERVATION OF DONOR LIVERS PRIOR
Manufacturer (Section D)
ORGANOX LIMITED
oxford science park, magdalen
centre, robert robinson avenue
oxford, oxfordshire, gb OX4 4 GA
UK  OX4 4GA
Manufacturer (Section G)
RAUMEDIC - MUENCHBERG
hermann-staudinger strasse 2
helmbrechts, bavaria, de 95233
GM   95233
Manufacturer Contact
ruth colwill
oxford science park, magdalen
centre, robert robinson avenue
oxford, oxfordshire, gb OX4 4-GA
UK   OX4 4GA
MDR Report Key19043401
Report Number3011560054-2024-00014
Device Sequence Number15264524
Product Code QQK
UDI-Device Identifier05060462240029
UDI-Public5060462240029
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P200035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD0146
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Initial Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-