| Model Number |
3799 |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient lost therapy due to ipg being inoperable.Reportedly, the patient went extended period of time without charging the ipg, rendering the ipg inoperable.Troubleshooting was unable to resolve the issue.As such, surgical intervention took place wherein the ipg was explanted and replaced with a new ipg to address the issue.Reportedly, therapy was confirmed post op.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient reported failing to charge their implant properly to abbott.The device has not been returned for analysis.A review of documentation supplied with the implant states that the implant must be charged every 30 days to avoid battery depletion.Based on the information received, the cause of the reported incident is consistent with user error.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h3.
|
| |
|