| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Halo (2227); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A4-a6: unk.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim #(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2 implantable collamer lens of diopter -11.5/1.5/164 (sphere/cylinder/axis) into the patient's left eye (os) on (b)(6) 2024.On (b)(6) 2024 the lens was explanted due to excessive vault, significant reduction of iridocorneal angles, and glare/haloes.On (b)(6) 2024 a replacement lens of a different model/ shorter length was implanted.Reportedly, "vault height satisfactory after icl exchange." the problem was resolved.The cause of the event was reported as other due to 'vault height too high.Icl oversized.'.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 - model #: vticm 13.2 corrected to vticm5_13.2.Claim #(b)(4).
|
| |
|
Search Alerts/Recalls
|