| Catalog Number |
320-42-13 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
02/16/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, the 69-year-old white male patient had a right tsa on (b)(6) 2020.The patient presented with disassociation of polyethylene on (b)(6)2023.Patient reported pulling on a 4-wheeler cable and injured shoulder.This event is related to (b)(4).The outcome of this event is considered resolved by the action of revision on (b)(6) 2023.The case report form indicates that this event is definitely not related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
|
| |
|
Additional Manufacturer Narrative
|
|
(d10) concomitant device(s): 320-20-00 - eq reverse torque defining screw kit : 6488539.320-10-05 - equinoxe reverse tray adapter plate tray +5 : 6511944.320-02-42 - rs expanded glenosphere 42mm, +4mm offset: 6574425.320-15-05 - eq rev locking screw : 6628478.Cs-10cc - intersep calcium sulfate : 016130.Cs-10cc - intersep calcium sulfate : 016372.(h3) pending evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Date of event is unknown.The revision reported cannot be confirmed from the information provided but may be the result of disassembly.However, the reported disassembly could not be confirmed.Potential contributions of user and patient-related considerations to the event, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|