MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-42-13
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 02/16/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported by the equinoxe shoulder study, the 69-year-old white male patient had a right tsa on (b)(6) 2020.The patient presented with disassociation of polyethylene on (b)(6)2023.Patient reported pulling on a 4-wheeler cable and injured shoulder.This event is related to (b)(4).The outcome of this event is considered resolved by the action of revision on (b)(6) 2023.The case report form indicates that this event is definitely not related to the device and/or to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
 
Additional Manufacturer Narrative
(d10) concomitant device(s): 320-20-00 - eq reverse torque defining screw kit : 6488539.320-10-05 - equinoxe reverse tray adapter plate tray +5 : 6511944.320-02-42 - rs expanded glenosphere 42mm, +4mm offset: 6574425.320-15-05 - eq rev locking screw : 6628478.Cs-10cc - intersep calcium sulfate : 016130.Cs-10cc - intersep calcium sulfate : 016372.(h3) pending evaluation.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Date of event is unknown.The revision reported cannot be confirmed from the information provided but may be the result of disassembly.However, the reported disassembly could not be confirmed.Potential contributions of user and patient-related considerations to the event, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REVERSE 42MM HUMERAL CONST LINER +2.5
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
matt collins
2320 nw 66th ct.
gainesville, FL 32653
3523771140
MDR Report Key19048182
Report Number1038671-2024-00746
Device Sequence Number20950328
Product Code KWT
UDI-Device Identifier10885862086723
UDI-Public10885862086723
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/29/2025
Device Catalogue Number320-42-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer05/06/2025
Initial Report FDA Received Date04/04/2024
Supplement Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Patient Age69 YR
Patient SexMale
Patient Weight100 KG
Patient RaceWhite
-
-