MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 55840006545
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D4: product identifiers are unknown e: facility details are unknown g3: pma / 510(k) # unknown medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from a healthcare professional via a manufacturer representative regarding a device used in an unknown spinal therapy.It was reported that the patient fell forward at the top of the construct and ended up pulling out the upper screws.There were no further symptoms or complications reported regarding the event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received from the manufacturer representative that no fall or other traumatic event occurred.At some point after surgery the patient¿s spine above the levels of fusion ¿fell forward¿, ie ¿shifted forward¿ ie went from a state of more lordosis to less lordosis.The increased strain on the fusion construct pulled out the screws at l2.No failure or breakage of any hardware in the construct.Patient had no symptoms indication the l2 screws pulled out and noticed on post-op x-ray follow-up.In second surgery the surgeon replaced the screws at l2 and extended the construct to l1.No adverse patient outcome after either surgery.
 
Additional Manufacturer Narrative
Additional information: h11 radiographic image review: 1.Intra-op fluoro l-s- i fusion lateral view.Interbody graft l3-4.L3 connects to construct at screws/hooks.L4-5 appears to be a separate construct.2.Lateral x-ray shows pull out of both l3 screws.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Brand name of the product updated - product code updated - product id updated - implanted date, - explanted date - 510(k) updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key19048351
Report Number1030489-2024-00246
Device Sequence Number7399929
Product Code KWP
UDI-Device Identifier00613994971753
UDI-Public00613994971753
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 07/01/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55840006545
Device Catalogue Number55840006545
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2024
Supplement Date Received by Manufacturer04/09/2024
04/09/2024
05/07/2024
06/06/2024
Initial Report FDA Received Date04/05/2024
Supplement Report FDA Received Date05/07/2024
05/07/2024
05/31/2024
07/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexUnknown
-
-