| Model Number |
RONDO 2 AUDIO PROCESSOR [BUTTON PROCESSOR 2] |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device was received at hq with a broken case part and a leaking battery.More information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The device was received at hq with a broken case part and a leaking battery.The user was reportedly not harmed by the leaking battery.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: during the repair process of a rondo 2 audio processor, a leaking battery was detected.The electrolyte corroded the silicone sleeve of the battery and oxidized a few parts.The damage to the coil and the rechargeable battery inside was most likely caused by strong mechanical stress.There have been no reports of patient injury from contact with leaking electrolyte.This is a final report.
|
| |
|