| Model Number |
CCWET3 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced blurred vision.Clinical reason was mentioned as lens exchange to help with distance.The iol was exchanged for another lens model following the initial implant procedure.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the 21.0 diopter, (1.5 extended depth of focus) lens was replaced with an unspecified atiol model.The product has not been received to evaluate.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to request for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|