DEPUY SPINE INC REDUCER KERRISON; MISC ORTHO SURGICAL INSTR Back to Search Results
Catalog Number 202000140
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2024 the instrument froze and would not allow the driver to move forward or backward.The instrument was removed before case started.There was no patient involvement.This report involves one (1) symphony oct system reducer kerrison.This is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: e3: reporter is a j&j employee.H3, h6: the subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: a review of the receiving inspection (ri) for reducer kerrison, was conducted identifying that lot number mi118521 was released in one batch.Batch1: lot qty of 34 units were released on nov 2, 2022 with no discrepancies.Supplier : depuy : micropulse incorporated.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the reducer kerrison shows signs of normal use.A dimensional inspection was not performed due to it is not applicable to the complaint condition.A functional test was performed, and the outer tube assembly does not slide into the inner tube assembly confirming the jammed condition.The stucked together can be due to internal damage however as it is not possible to disassemble a potential cause cannot be established as part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the device would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCER KERRISON
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19049378
Report Number1526439-2024-01226
Device Sequence Number7355287
Product Code LXH
UDI-Device Identifier10705034530567
UDI-Public10705034530567
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number202000140
Device Lot NumberMI118521
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2024
Supplement Date Received by Manufacturer04/15/2024
05/21/2024
Initial Report FDA Received Date04/05/2024
Supplement Report FDA Received Date05/09/2024
05/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REDUCER KERRISON
Patient SexUnknown
-
-