• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DIREXION; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DIREXION; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 80618
Medical Device Problem Codes Defective Device (2588); Difficult to Open or Remove Packaging Material (2922); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
Reportable upon analysis completed 21march2024.It was reported that this microcatheter had kinked.During imaging of a renal embolization procedure using a direxion microcatheter, it was found that the device was faulty during tracking though the anatomy and the hub had kinked.It was noted that the carrier tube was hydrated prior to the removal of the catheter; however, there was a little resistance removing the catheter from the carrier tube.A new device was used to complete the procedure, and no patient complications were reported.However, device analysis revealed a fracture and a device separation.
 
Additional Manufacturer Narrative
Device eval by manufacturer: this direxion microcatheter was returned and analyzed.Visual inspection of the complaint device revealed a fracture located 3.5 cm, and a separation of the shaft at 24.5 cm from the microcatheter hub.Functional testing was not able to be performed due to the separation.No other visual abnormalities were noted.Analysis of the device found the nitinol shaft was fractured and separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key19049957
Report Number2124215-2024-20287
Device Sequence Number8482767
Product Code KRA
UDI-Device Identifier08714729839453
UDI-Public08714729839453
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2024
Device Model Number80618
Device Catalogue Number80618
Device Lot Number0029394184
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2024
Initial Report FDA Received Date04/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-