| Model Number |
80618 |
| Medical Device Problem Codes |
Defective Device (2588); Difficult to Open or Remove Packaging Material (2922); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable upon analysis completed 21march2024.It was reported that this microcatheter had kinked.During imaging of a renal embolization procedure using a direxion microcatheter, it was found that the device was faulty during tracking though the anatomy and the hub had kinked.It was noted that the carrier tube was hydrated prior to the removal of the catheter; however, there was a little resistance removing the catheter from the carrier tube.A new device was used to complete the procedure, and no patient complications were reported.However, device analysis revealed a fracture and a device separation.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by manufacturer: this direxion microcatheter was returned and analyzed.Visual inspection of the complaint device revealed a fracture located 3.5 cm, and a separation of the shaft at 24.5 cm from the microcatheter hub.Functional testing was not able to be performed due to the separation.No other visual abnormalities were noted.Analysis of the device found the nitinol shaft was fractured and separated.
|
| |
|