| Catalog Number |
528135 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the staple will not come out from the stapler.A new stapler was used to complete the procedure.No patient harm or injury.The patient status is reported as "normal".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of failure mode reported in the current manufacturing process was conducted as follows: 20 staplers were taken from the current production from part number 528135 visistat 35r 6/box lot number 73c2400176 the staplers were functionally inspected (fired) and issue reported "misfire/jam - staples not releasing" was not observed in the current manufacturing process, the staples were loaded and released correctly.Dhr investigation did not show issues related to complaint.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that the staple will not come out from the stapler.A new stapler was used to complete the procedure.No patient harm or injury.The patient status is reported as "normal".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint of misfire/jam - staples not releasing was confirmed based upon the sample received.Visual examination revealed that the first staple appeared to be misaligned.It was observed that the first misaligned staple was positioned upside-down in the cover block.Upon functional inspection, the misalignment of the staples prevented the remaining staples from consistently firing correctly.The root cause of the staple misalignment could not be conclusively determined.A device history record review was performed on the device with no evidence to suggest a manufacturing related cause.An investigation within teleflex was initiated to further evaluate this complaint issue.Teleflex will continue to monitor and trend on complaints of this nature.
|
| |
|