| Model Number |
CF-HQ190I |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device inspection, the colonovideoscope's nozzle was clogged and the air and water cylinders had foreign objects.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: e1.Additional information added to field h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.The legal manufacture reviewed the customer provided the cleaning disinfection and sterilization processes where no obvious deviations from instructions for use were identified.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.The event can be detected/prevented by following the instructions for use: instructions for use states that detection method in gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection.Instructions for use states that preventive measure in gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|