| Model Number |
M0063903100 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Imdrf device code a0414 captures the reportable event of sheath torn.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a nitinol retrieval coil device was used during a ureteroscopy procedure performed on (b)(6) 2024.During the procedure, it was found that the sheath was torn upon opening the package.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0414 captures the reportable event of sheath torn.H2: additional information block b5 (describe event or problem) block d7a (sud reprocessed and reused) block h8 (usage of device).
|
| |
|
Event or Problem Description
|
|
It was reported that a nitinol retrieval coil device was used during a ureteroscopy procedure.During the procedure, it was found that the sheath was torn at the proximal section close to the mid of the sheath.The procedure was completed with another of the same device and there were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0414 captures the reportable event of sheath torn.Block h11: investigation results: the returned stone cone was analyzed, and a visual evaluation noted that the coil was received in an open state.It was also noted that the blue sheath was broken exposing the white sheath.Functional examination was performed, and the coil was able to fully open and close even though the blue sheath was broken.A labeling review was performed and, there is no evidence that this device was not used in accordance with instructions for use.With all the available information, boston scientific concludes that the reported event was confirmed.Based on the investigation it is most likely that the damage was caused during the test of the device with the use of excessive force or manipulation.When force was exerted when trying to open/close the coil, the blue sheath broke.Additionally, the ifu states, if resistance is encountered while attempting to withdraw the coil do not exert excessive force.To release the object from the device, advance the sheath to straighten the coil and remove the device.Therefore, the most probable root cause of unintended use error caused or contributed to event was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that a nitinol retrieval coil device was used during a ureteroscopy procedure.During the procedure, it was found that the sheath was torn at the proximal section close to the mid of the sheath.The procedure was completed with another of the same device and there were no patient complications reported as a result of this event.
|
| |
|